Help with sigma metric analysis

dsc_0900edp

I’ve been interested in glucose meter specifications and evaluations. There are three glucose meter specifications sources:

FDA glucose meter guidance
ISO 15197:2013
glucose meter error grids

There are various ways to evaluate glucose meter performance. What I wished to look at was the combination of sigma metric analysis and the error grid. I found this article about the sigma metric analysis and glucose meters.

After looking at this, I understand how to construct these so-called method decision charts (MEDX). But here’s my problem. In these charts, the total allowable error TEa is a constant – this is not the case for TEa for error grids. The TEa changes with the glucose concentration. Moreover, it is not even the same at a specific glucose concentration because the “A” zone limits of an error grid (I’m using the Parkes error grid) are not symmetrical.

I have simulated data with a fixed bias and constant CV throughout the glucose meter range. But with a changing TEa, the estimated sigma also changes with glucose concentration.

So I’m not sure how to proceed.