Six Sigma can be dangerous to your health

sigma

At a recent conference, there were several presentations about six sigma for clinical laboratory assays. To recall, sigma is calculated as Sigma = (TEa – bias)/CV where

TEa is the total allowable error
Bias is the inaccuracy of the measurement procedure
CV is the imprecision of the measurement procedure

The problem with six sigma is that’s it taken as a sole measure of quality – that is, if you have a high sigma value (greater than 6) then your assay is assured of high quality. The rest of this entry explains why this is wrong.

First, TEa (total allowable error) is often specially called out as medically acceptable limits. One need only read the ISO 15197 standard for glucose to see this connection. I have previously commented about this standard. The implied meaning of medically acceptable limits in shown in below.

figure 1

This is simply not the real world. Taguchi long ago specified a more realistic quadratic model of worth, which is shown below, superimposed on the original figure but in green.

figure 2

Thus points A and B are similar in bias and are similar in causing (or not causing) medically unacceptable results. It is also likely then that if point A is ok, then so is point B. It is only when one gets far away from these limits that one is almost certain to have results that can cause harm. This is shown below with point C.

figure 3

This can also be expressed as an error grid such as those for glucose. So the “sigma” calculations really only express the zone A region (grey) where 95% or more of the results should be. Zone B (white) can contain up to 5% of the results and zone C (dark grey) should contain no results. The error grid contains more information since each set of limits is different for each concentration. An error grid is shown below, taken from FDA guidance. In the guidance, WM is the test method and CM is the reference method. (In the document WM=waiver method and CM=comparative method).

figure 4

So the problem is that sigma only accounts for zone A, but patients are harmed by values in zone C!

Now one might argue that there is nevertheless a relationship between sigma and the three zones, meaning that high sigma values are unlikely to have values in zone C and low sigma values are likely to have such values. This is also not true. Here is why.

1.       Often incorrect models are used to asses total error – see here.

2.       In estimating bias and CV, outliers – the very values that cause harm - are often thrown out.

3.       All sigma calculations are based on the assumption that the data are normally distributed. Most data do not fulfill this criterion. This means that often there are more frequent values in the tails of the distribution (again, this is zone C) than expected by calculations based on the normal distribution

4.       And maybe the biggest reason of all, values can occur in zone C that have nothing to do with the analytical process. If there is a patient sample mix-up, this can occur and these values are excluded (when detected) from virtually all analytical evaluations.

Think of it this way. If a loved one suffered medical harm, due in part to an erroneous lab result, would it make you feel better to know that the assay had a high sigma value? And would you associate that assay with quality?

I will comment on how one can address these issues in a future entry.

Leave a Reply